Midmark Error Codes

C010: POWER MODE SYSTEM PWR LOSS, ITEMS NOT STERILE PUSH STOP TO RESTART: 

  • Secure power cord connections. At wall outlet and at sterilizer receptacle.
  • Unplug unit for 60 seconds, then plug unit into a dedicated, properly rated circuit.

  • Secure / Repair loose or damaged connections.
  • Replace power cord/ receptacle.
  • Secure/replace ribbon harness.
  • Replace Main PC board.

C102: FILL MODE STOP PRESSED 

  • Press STOP button and initiate a new cycle.
  • Replace touch pad.

C645: VENT MODE PRESSURE HIGH, ITEMS NOT STERILE PUSH STOP TO RESTART
Absolute pressure inside chamber is greater than 0.7 psig (5 kPag) for longer than 10 minutes. i.e. Vent mode is not complete.

  • Perform monthly maintenance.
  • Clean / Replace filter. Clean with Speed-Clean and brush.

  • Clean / Replace vent valve.
  • Replace Main PC board.
  • Replace Thermal Regulator Wax Motor or Thermal Sensor.
  • Install the tank higher than the facility drain.